Torsional mode phacoemulsification: Effective, safe cataract surgery technique of the future


1 Department of Ophthalmology, Ain-Shams University, Cairo, Egypt and Ain Shams University,Cairo, Egypt; Consultant cataract and glaucoma, Mouwasat Hospital, Kingdom of Saudi Arabia
2 Department of Ophthalmology, Ain-Shams University, Cairo, Egypt

Date of Web Publication22-Mar-2010

Correspondence Address:
Ahmed M El-Moatassem Kotb
46 Ahmed Mohamed Ibrahim Street, Nasr City Area, Cairo 11471, Egypt

Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0974-9233.61220

Rights and Permissions
   Abstract 

Purpose: To compare various outcome measures using torsional mode and longitudinal mode in the phacoemulsification of cataract with different nuclear densities.
Setting: Magrabi Eye Hospitals, Kingdom of Saudi Arabia.
Design: A randomized comparative clinical study.
Materials and Methods: This study includes 200 eyes of 156 patients (100 in the ultrasound longitudinal "US" group and 100 in the torsional group). All eyes received AcrySof® single piece intraocular lens (Alcon Surgical, Fort Worth, TX). The primary outcome measures were ultrasound time (UST), cumulative dissipated energy (CDE), and surgical complications. Postoperative outcome measures were the degree of corneal edema on the first postoperative day and final best corrected visual acuity (BCVA) and CCT (central corneal thickness).
Results: The differences in UST and CDE between subgroups of nucleus hardness were statistically significant (P < 0.01). The UST and CDE consistently increased in eyes with higher grades of nucleus density. On day one, the mean BCVA was 0.61 ± 0.13 decimals in the ultrasound (US) group and 0.67 ± 0.11 decimals in the torsional group (significant P < 0.05).Corneal edema was significantly less in the torsional group (P < 0.05). At 30 days, the mean BCVA was 0.94 ± 0.22 decimals in the US group and 1.0 ± 0.12 decimals in the torsional group but this difference was not statistically different (P > 0.05).
Conclusions: The torsional mode provides an effective and safe method for cataract removal with lower energy usage as compared to longitudinal traditional phacoemulsification. However, the final visual outcome was similar for both study groups.

Keywords: Cataract Densities, Longitudinal Mode Phacoemulsification, Torsional Mode


How to cite this article:
El-Moatassem Kotb AM, Gamil MM. Torsional mode phacoemulsification: Effective, safe cataract surgery technique of the future. Middle East Afr J Ophthalmol 2010;17:69-73

How to cite this URL:
El-Moatassem Kotb AM, Gamil MM. Torsional mode phacoemulsification: Effective, safe cataract surgery technique of the future. Middle East Afr J Ophthalmol [serial online] 2010 [cited 2019 Jun 26];17:69-73. Available from: http://www.meajo.org/text.asp?2010/17/1/69/61220


   Introduction Top


Advancement in phacoemulsification technology allows use of smaller surgical instruments, flexible intraocular lenses (IOL) and advanced management software for phacoemulsification units further reducing incision size and tissue trauma and promoting faster functional recovery. In conventional US mode, the US power to emulsify the lens is derived from the longitudinal excursion of the phacoemulsification needle. [1] Phacoemulsification tip moves forward and backward at a high frequency. However, in the conventional US mode, phacoemulsification can produce repulsion effect because the phaco tip pushes the nucleus away with each stroke as it moves forward. Thus, the US is interrupted and efficiency of phacoemulsification is compromised. [2]

In January 2006, Alcon Surgical incorporated OZil torsional into the Infiniti Vision System. The OZil torsional portion is a hardware and software upgrade of the machine and includes a dedicated hand-piece that produces rotary oscillations of the phacoemulsification tip with a frequency of 32 KHz. It is suggested that the OZil torsional oscillation effect and the incorporated improvements reduce the amount of phacoemulsification energy and increase the efficiency required to remove the cataractous nucleus because it does not produce repulsion and breaks up cataract by shearing and not by conventional jackhammer effect. [2]

Improvements in phacoemulsification surgery afforded by torsional phacoemulsification are found in the OZil hand-piece's unique movement profile and reduced frequency. [3] The shaft of the torsional hand-piece oscillates, and the oscillatory motion is translated to side-to-side movement at the tip. Torsional phacoemulsification uses angulated tips, so the motion of the tip within the nuclear tissue is up to three times greater than movement of the tip's shaft, according to Dr. Mackool. [3]

We designed this study to compare different clinical outcomes using torsional versus longitudinal phacoemulsification mode in different cataract densities.


   Materials And Methods Top


Participants

Patients aged 54 years or older, with the diagnosis of age-related cataract, have been included in the study. Patients who had other ocular or systemic disorders affecting vision were excluded. These included patients with diabetic retinopathy, glaucoma, age-related macular degeneration, uveitis, corneal endothelial disease or previous intraocular surgery.

Methods and surgical technique

First, patients are enrolled and informed consent is obtained from them. Standard preoperative ophthalmological examinations including lens nucleus density grading according to the Lens OpacitiesClassification System II (LOCS II) [4] are then performed.

Surgical technique

All patients received periocular anesthesia [5] and surgeries were performed by the same surgeon (A.K.). A 2.8 mm self sealing limbal incision was made at 12 O'clock and a paracentesis using 15º blades performed at about 60 0 from the main incision. Sodium hyaluronate 3.0%- chondroitin sulfate 4.0% (Viscoat, Alcon Surgical, Fort Worth, TX) was used to reform and stabilize the surgical planes and protect the corneal endothelium. A 5.5 to 6.0 mm continuous curvilinear capsulorhexis was performed with a bent 27-gauge needle.

A standard horizontal chopping technique [6] was used with the Infiniti Vision System with the US pulse mode or torsional mode. We used the OZil torsional hand-piece through a 2.8 mm incision (the 0.9 mm tapered Kelman microtip with a 30º bevel [Alcon Laboratories, Inc. Fort Worth, TX]).

The surgical parameters used in either group are summarized in [Table 1]. An intraocular lens AcrySof® Single Piece Intraocular Lens (Alcon Surgical, Fort Worth, TX) was inserted with the injector system through the 2.8 mm incision into the bag. The limbal wound was not sutured [Figure 1].

The main outcome parameters were mean UST and mean cumulative dissipated energy (CDE). The UST represents the number of seconds the foot pedal remained in the third position. The mean CDE power indicates the mean percentage of power spent during the UST. (CDE = mean US power x UST)The UST and CDE values in torsional and phaco modes were automatically calculated by the device and displayed on the monitor of the phaco machine. In torsional mode, the CDE was calculated as follows: Torsional amplitude x torsional time x 0.4. The frequency of the phaco tip in torsional mode was 80% of that in standard phacoemulsification (32 Khz in torsional versus 40 Khz in standard phacoemulsification), and the stroke distance of the phaco tip in torsional mode was half that in standard phacoemulsification. This justified the coefficient of 0.4.

The patients were seen one day, seven days and 30 days postoperatively. The postoperative best corrected visual acuity (BCVA), degree of corneal edema and intra and post-operative complications were all documented. Corneal edema was graded as trace; where there was minimal corneal clouding and thickening in relation to incision sites, mild - corneal clouding and thickening affecting less than 25% of the cornea with no Descemet's folds and clear iris details, moderate-corneal clouding and thickening affecting more than 25% of the cornea with few Descemet's folds and hazy iris details and severe-corneal clouding and thickening affecting more than 50% of the cornea with more Descemet's folds and no view of iris details. Central corneal thickness was measured with Visante OCT 1000 (Carl Zeiss Meditec) anterior segment imaging.

The SPSS 13.0 (SPSS Inc. Chicago, Illinois) was used. The independent- samples t test, was used to compare decimal visual acuity, UST, CDE and CCT. Chi-square test was used to compare qualitative variables while the unpaired t-test was used to compare quantitative variables between groups. Test was considered significant (S) if P < 0.05, highly significant (HS) if P < 0.01 and non significant (NS) if P > 0.05. [7]


   Results Top


A total of 200 eyes were enrolled in this study (156 patients); 100 in the conventional US group and 100 in the torsional US group according to LOCS II classification. The mean age was 67.9 years ± 3.9.

The differences in UST and CDE between subgroups of nucleus hardness were statistically significant ( P < 0.01). The UST and CDE consistently increased in eyes with higher grades of nucleus density [Table 2] and [Table 3]. The mean UST and CDE were lower in the torsional group than in the US group for all grades of nucleus densities. The difference was significant for Grade-1 and Grade-4. [Table 2],[Table 3] and [Figure 2] and [Figure 3] illustrate the differences in the UST and CDE between the two groups at different nucleus densities.

On day one, the mean BCVA for all grades was 0.61 ± 0.13 decimals in the US group and 0.67 ± 0.11 decimals in the torsional group. This difference was statistically significant ( P < 0.05).

By 30 days, the mean BCVA for all grades was 0.94 ± 0.22 decimals in the US group and 1.0 ± 0.12 decimals in the torsional group the difference was not statistically significant. ( P > 0.05).

The frequency of severe central corneal edema and Descemet's striae was higher in the US group than in the torsional group [Table 4]. On day one, the difference between the two groups was statistically significant ( P < 0.05). By 30 days, there was no corneal edema in either group. Except for corneal edema, no intraoperative or postoperative complications were noted in either group.


   Discussion Top


In recent years, technological developments have led to advances in cataract surgery and new methods to reduce US energy have been achieved. Examples include non ultrasonic energy such as sonic frequencies, NeoSoniX-generated tip rotation, [8],[9],[10] and pulse water-jet technology. Restricting energy to fractions of second pulses or bursts, [11],[12] and millisecond-level microburst, [13] tip design, [14] and specific chopping technique [15],[16],[17],[18],[19],[20] are favorable to lower energy. For the lens to be removed efficiently and safely, the risk of ultrasound induced endothelial cell loss should be minimized. Reducing phacoemulsification energy and time are the main objectives of future improvement.

A recent study by Davison compared surgeon-generated tip travel and surgical time of longitudinal and torsional phacoemulsification with straight and angled tips. He found that the angled tip/torsional US combination reduced tip travel by more than 40% compared with the straight tip/longitudinal US. He stated that shorter cumulative tip travel and less procedure time imply increased nuclear followability, fewer reacquisition movements, and increased phacoemulsification efficiency and safety. [21]

In this study we compared the safety and efficacy of torsional mode with longitudinal phacoemulsification in different grades of nucleus densities. Our results demonstrate that torsional phacoemulsification produces a safe and efficient mode of phacoemulsification than longitudinal mode with reduced mean UST and CDE in all grades of nucleus densities. Our results are supported by results obtained by Yizhi Liu et al. [2] Energy saving (as determined by CDE) was significant in nucleus Grade-1 and Grade-4 following torsional phacoemulsification. This was reflected in the absence or traces of corneal edema in most patients on the first day and only a few patients with mild or moderate corneal edema that resolved within one week of adequate treatment following torsional phacoemulsification.

As mentioned before, torsional phacoemulsification causes less repulsion of nuclear materials than conventional phacoemulsification, so lens materials stays at the tip most of the time. As a consequence, in our study less vacuum level was used with torsional mode compared to the longitudinal mode, this counteracts surge and the outcome is a safer surgery.

The mean BCVA at day one was significantly better in the torsional group which was attributable to less corneal edema secondary to less UST and CDE. However, at 30 days postoperative, the mean BCVA was nearly the same in both groups with no statistically significant differences. This emphasize that torsional phacoemulsification has better visual outcomes in the early postoperative period.

One of the limitations of this study is that we did not evaluate the effect of surgery on the corneal endothelial cell count. The corneal thickness measurement by anterior segment OCT was used as a surrogate for endothelial cell function thus indirectly evaluating the effect of surgery on corneal endothelium. Further, other studies have demonstrated clinically significant decrease in endothelial cell loss with torsional phacoemulsification mode in comparison to conventional phacoemulsification. [2] The loss of endothelial cell count also correlated with the size of incision, where 2.2 mm incision with phacomeulsification sleeve (Ultrasleeve TM ) produced clinically significant less endothelial cell loss than standard 2.8 mm incision with microsleeve. [22]


   Conclusion Top


Torsional mode phacoemulsification is an effective and safe technique for cataract surgery. It has the advantage of reducing UST, as well as effective energy used with earlier visual rehabilitation that makes torsional mode phacoemulsification the technique of the future.



 
   References Top

1.Packer M, Fishkind WJ, Fine IH, Seibel BS, Hoffman RS. The physics of phaco: A review. J Cataract Refract Surg 2005;31:424-31.  Back to cited text no. 1      
2.Liu Y, Zeng M, Liu X, Luo L, Yuan Z, Xia Y, et al. Torsional mode versus conventional ultrasound mode phacoemulsification Randomized comparative clinical study. J Cataract Refract Surg 2007;33:287-92.  Back to cited text no. 2      
3.Mackool R. Torsional phacoemulsification: The elimination of lens chatter and thermal energy. Presented as an Alcon booth presentation at: Annual Meeting of the American Society of Cataract and Refractive Surgery; March 17-22,2006; San Francisco.   Back to cited text no. 3      
4.Chylack LT Jr, Leske MC, McCarthy D, Khu P, Kashiwagi T, Sperduto R. Lens Opacities Classification System II (LOCS II). Arch Ophthalmol 1989;107:991-7.  Back to cited text no. 4      
5.Bloomberg LB: Anterior periocular anaesthesia: Five years experience. J Cataract Refract Surg 1991; 17: 508-11.  Back to cited text no. 5      
6.Chang DF. Converting to Phaco Chop: Why? Which technique? How? Ophthalmic Practice 1999;17:202-10.  Back to cited text no. 6      
7.Miller MC, Knapp RG. Clinical epidemiology and biostatistics, published by Williams and Wilkins: Maryland: 3 rd edition 1992.  Back to cited text no. 7      
8.Hoffman RS, Fine IH, Packer M, Brown LK. Comparison of sonic and ultrasonic phacoemulsification using the Staar Sonic Wave system. J Cataract Refract Surg 2002;28:1581-4.  Back to cited text no. 8      
9.Davison JA. Ultrasonic power reduction during phacoemulsification using adjunctive NeoSoniX technology. J Cataract Refract Surg 2005;31:1015-9.  Back to cited text no. 9      
10.Vasavada AR, Raj SM, Lee YC. NeoSoniX ultrasound versus ultrasound alone for phacoemulsification; randomized clinical trial. J Cataract Refract Surg 2004;30:2332-5.  Back to cited text no. 10      
11.Fine IH, Packer M, Hoffman RS. Power modulations in new phacoemulsification technology: Improved outcomes. J Cataract Refract Surg 2004;30:1014-9.  Back to cited text no. 11      
12.Badoza D, Fernández Mendy J, Ganly M. Phacoemulsification using the burst mode. J Cataract Refract Surg 2003;29:1101-5.  Back to cited text no. 12      
13.Rose AD, Kanade V. Thermal imaging study comparing phacoemulsification with the Sovereign WhiteStar system to the Legacy with AdvanTec and NeoSoniX system. Am J Ophthalmol 2006;141:322-6.  Back to cited text no. 13      
14.McNeill JI. Flared phacoemulsification tips to decrease ultrasound time and energy in cataract surgery. J Cataract Refract Surg 2001;27:1433-6.  Back to cited text no. 14      
15.Vergés C, Llevat E. Laser cataract surgery: Technique and clinical results. J Cataract Refract Surg 2003;29:1339-45.  Back to cited text no. 15      
16.Güell JL, Vázquez M, Lucena J, Velasco F, Manero F. Phaco rolling technique. J Cataract Refract Surg 2004;30:2043-5.  Back to cited text no. 16      
17.Sebban I. Phaco one-chop nucleotomy. J Cataract Refract Surg 2002;28:1325-9.  Back to cited text no. 17      
18.Fine IH, Packer M, Hoffman RS. Use of power modulations in phacoemulsification; choo-choo chop and flip phacoemulsification. J Cataract Refract Surg 2001;27:188-97.  Back to cited text no. 18      
19.Arshinoff SA. Phaco slice and separate. J Cataract Refract Surg 1999;25:474-8.  Back to cited text no. 19      
20.Wong T, Hingorani M, Lee V. Phacoemulsification time and power requirements in phaco chop and divide and conquer nucleofractis techniques. J Cataract Refract Surg 2000;26:1374-8.  Back to cited text no. 20      
21.Davison JA. Cumulative tip travel and implied followability of longitudinal and torsional phacoemulsification. J Cataract Refract Surg 2008;34:986-90.  Back to cited text no. 21      
22.Berdahl JP, Jun B, DeStafeno JJ, Kim T. Comparison of a torsional handpiece through microincision versus standard clear corneal cataract wounds. J Cataract Refract Surg 2008;2091-5.  Back to cited text no. 22      


    Figures

  [Figure 1], [Figure 2], [Figure 3]
 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
    Introduction
    Materials And Me...
    Results
    Discussion
    Conclusion
    References